EALTHCARE
5394152, Revision 5
D
IRECTION
8-2-25-3
Removal procedure (cont'd)
8.) Unplug the J14 connector, and unscrew 2 screws. Pull out the APS II Assy carefully.
Note: AC power input connector should be passed through the square hole on chassis bottom.
9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-26
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-9-6
Section 10-5-2
Section 10-5-5
8-2-25-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-104 Pull out the APS Assy
Functional Tests
Section
Functional Test / Diagnostic Test
APS II Assy PCB Function Validation
Functional Checks (See Also Chapter 4)
Power On/Boot Up
Power Off / Shutdown
Procedure
Physical Inspection
Chapter 8 -
LOGIQ™ P6/P6 P
S
RO
ERVICE
Debrief Script
"Service Manual, Direction
5394152, Rev 1+, Section 8-2-25. Equipment
passed all required tests and is ready for use. "
M
ANUAL
Page 8-65