Item
Reading Modes
Objectives
Fluorescence Color Channels
Temperature Control
Ambient Operating Conditions
Ambient Storage Conditions
Altitude Restrictions
IEC Installation Category
Pollution Degree
Data Connection
IEC Ingress Protection
42
Description
Fluorescence
Transmitted Light (Brightfield)
Colorimetric
4x
10x
20x (optional)
40x (optional)
63x (optional)
FITC
DAPI (optional)
TRITC (optional)
CY5 (optional)
6°C (11°F) above ambient up to 40°C (104°F)
Temperature: 18°C to 30°C (59°F to 86°F)
Temperature (recommended when using temperature
control): 20°C to 24°C (68°F to 75°F)
Relative Humidity: 20% to 75% (non-condensing)
Temperature: -20°C to +60°C (-4°F to 140°F)
Relative Humidity: 15% to 75% relative humidity (non-
condensing)
Up to 2000 m (6,562 ft)
II
2
Two (2) Ethernet ports
IP20
5063666 B