SMC Networks ISA3 Series Operation Manual page 79

Digital gap checker pressure switch
Hide thumbs Also See for ISA3 Series:
Table of Contents

Advertisement

: When returning to measurement mode is required during the setting, press the SET button for 2 seconds or longer. Finish the
calibration and return to measurement mode.
Step2: Subtle adjustment of the pressure display
value
Ensure that the pressure on the SUP port side and
OUT port side are the same.
Measure the applied pressure and press the UP or DOWN button to adjust
the value to be the same.
When adjustment rate is changed, the pressure value after the adjustment
will be displayed on the main display.
When the current pressure is unknown, press the SET button without
changing the adjustment rate while the SUP side pressure and OUT side
pressure are equalized using a plug. It is possible to reduce the error
between the SUP side and OUT side pressure.
: OUT port side pressure will be adjusted automatically.
: When the pressure after the adjustment is below the supply pressure lower limit (80 kPa) or
exceeds the display set range upper limit (220 kPa), or when the automatic adjustment of the
OUT port side pressure fails, Er30 is displayed to return to the display value adjustment.
: If the display value fine adjustment is not performed, press the UP and DOWN buttons for 1
second or longer.
The screen will move back to the calibration selection mode.
: The display value can be adjusted within 5% R.D.
Press the SET button to set.
Step3: Adjustment of detection distance
(guideline) display value
Adjust the nozzle gap to achieve applicable gap using
a master work piece or shim.
Press the UP or DOWN button to change the
adjustment rate.
The current detection distance is displayed on the main display.
When the adjustment rate is changed, the main display shows the value
after the adjustment, so change it to the value to be changed.
: The display value can be adjusted within 5% R.D.
Press the SET button to set.
[F95] Setting of calibration function is completed
-78-
4
Move on to adjustment of detection
distance (guideline) display value.
Return to function selection mode.
No.PS※※-OMW0009-A