6.2

Usage instructions for DS mode

6.2.1
Call-up of DS mode
Call-up is possible at all times during operation even with INOP.
There are two possibilities:
-
Call-up via keys:
Press both "Silence" and "negative temp." keys for four seconds during
operation.
-
Call-up without keys
Call-up is effected via a short-circuit produced for approximately 1 s for
example by way of a jumper on the controller PCB after the self-test.
Important: Error message after 30 s with sustained short circuit.
Position in normal operation without jumper or jumper in park position 2-3:
Call-up of service mode. Short circuit for max. 30 seconds between 1-2:
6132.200
Babytherm 8000
A B C
D E F
37°C
35°C
X1
X1
Repair Instructions
Horn
Check
off
X12
1
X12
1
04.01
Page 70 of 115