The following is required for conversion from 220/240V to 100/127V:
-
Accessory set, voltage 100/127V
Important: It is not possible to change the voltage for the optional height
adjustment!
4.2.3
Testing electronics, complete
Assemble unit ready for operation, but do not fix electronics assembly in position in
housing. Connector for attaching sensor cable to unit is therefore accessible.
Switch on unit:
-
All displays and horn are actuated for 1.2 s
-
Then:
o
o
o
-
Without prior pressing of range expansion key setpoint can be adjusted to
between 35.0 °C and 37.0 °C.
-
After prior pressing of range expansion key setpoint can be adjusted in each
case to min 30.0 °C or max. 38.5 °C.
-
Adjust setpoint to 37.0 °C.
-
Press "check" key:
o
o
o
-
If setpoint higher than actual value:
6132.200
Babytherm 8000
Setpoint display = 37.0 °C
Measured value and label "SEt" flash up alternately on actual value
display. Important: measuring range 5 °C - 45 °C.
Pressing one of the keys " ⇑ " or " ⇓ " setpoint produces constant
actual display.
All displays and horn are actuated for 2 s.
All displays and horn then off for 2 s.
This is followed by return to original display.