The heating can be switched on only in this mode for max. 60 s with the "check"
key. This causes the heating to be clocked (1 s on and 0.5 s off).
Heating test:
-
Switch on safety relay in mode b.
-
Switch on heating in mode C:
Alternating feedback E = 1 and 0 (in mode C)
Mode number can be switched with the keys "positive temp." or "negative temp.".
6.3.d
Mode d Position of mains switch
Readout of position of mains-switch auxiliary contacts for warm/cold-start
recognition.
F
A B C
D
6132.200
Babytherm 8000
=
HEA as flag for heater (heating)
=
Request, switching status, heating
0
=
1
=
=
Heating feedback
0
=
1
=
Back
A B C
D E F
37°C
35°C
=
Mode number d flashing
=
P F S as flag for power fail switch
=
Position of mains-switch auxiliary contacts
Repair Instructions
Heating off
Heating on
Heating off detected
Heating on detected
Horn
Check
off