Switch on unit
-
Create short circuit on test board between TP 3 (+ 5 VD) and TP 30 (horn
off).
-
Interrupt mains connection but do not switch off unit.
-
Reestablish mains connection
-
Additionally create short circuit on test board between TP 3 (+ 5 VD) and TP
35 (horn on ).
4.4.3.4
Testing visual mains failure alarm
Prerequisite: mains-switch auxiliary contacts OK.
-
Assemble unit ready for operation
-
Create mains connection
-
Switch on unit
-
Interrupt mains connection
4.4.4
Replacement in repair situation
The following spare parts are available for repairs:
-
Power PCB, complete (see TSB # 4 or IDM # 5)
o
o
6132.200
Babytherm 8000
Horn on
Horn off
Horn on (delay time due to monoflop)
Horn off
Horn on
Visual and acoustic mains failure alarm
Relay 5V / 8A
Fuse T 800 mA
Repair Instructions
82 90 771
18 37 001
18 15 075