Health Status and Alarm
Management

Health Status Overview

Selecting the Health Status (green heart) icon in the home
screen displays the following screen:
Figure 65. Health Status Screen
Door Openings: Total displays the number of times the
door has been opened since the last reset. Last Open
displays the duration the door was last opened. Press the
Reset icon to change the Door Openings values to 0.
Temperature Excursions: Actual displays the current
cabinet temperature. Warm and Cold displays are the
highest and lowest temperatures recorded since the last
reset. Press the Reset icon to set the Warm and Cold
values to the current cabinet temperature.
Conditions: Voltage displays the input voltage. Ambient
displays the unit ambient temperature. By pressing the
Conditions Detail button the following screen will be
displayed:
Figure 66. Conditions Detail Screen
Ultra Low Temperature Freezers
Power Modes: Displays the state of the on-board
voltage conditioning system.
Main Battery: Displays the voltage of the main battery.
There is also a 12 month countdown timer that should be
reset every time the battery is replaced.
Backup System (BUS): The BUS information will be
displayed if a BUS is present.
Selecting the System Dashboard button in Figure 67 
displays the on board sensor readings.
Figure 67. System Dashboard Screen
When a BUS is present, the BUS RTD and BUS battery
values are also displayed.

Health Status and Alarm Management |

29