Install Wall Bumpers

Remote Alarm Contacts

Thermo Fisher Scientific
The parts bag, located inside the cabinet, contains the following parts.
Quantity
Stock #
2
510016
2
380520
Install the bolts into the pre-tapped holes on the back of the freezer. Install
a neoprene cap on each bolt. Refer to Figures 1-3 and 1-4 for the locations
of the pre-tapped holes.
See Figures 1-3 and 1-4 for the location of the remote alarm contacts. The
remote alarm connector is located in the parts bag provided with the
manual. It must be installed if connecting the freezer to an alarm system.
After installing the wiring from the alarm system to the connector, install
the connector to the freezer microboard and secure with the two screws
provided. The remote alarm provides a NO (normally open) output, a NC
normally closed) output and COM (common). The contacts will trip on a
power outage, high temperature alarm or low temperature alarm. Figure
1-10 shows the remote contacts in alarm state.
Figure 1-10. Remote Alarm Contacts
Description
1/4-20 x 5-1/2" Bolt
Neoprene Cap
Section 1
Installation and Start-up
Purpose
Wall Bumper
Cap Protector
7010708