Controls
Temperature setpoints can be set in the Controls screen. 
To select a temperature setpoint,
Select the Controls button to navigate to the Controls screen.
Figure 35. Controls Screen
Operating Mode: The default setting is the Standard
mode which provides a balance between power
consumption and peak variation performance. 
High Performance mode provides minimum temperature
peak variation.
Note: Customers performing on-site temperature calibration
may observe as much as a 2°C variation when an external
probe is placed next to the freezer control probe. 
This variation is due to optimization of the control system to
ensure temperature uniformity throughout the chamber.
Power Recovery Delay: Press this button to set the
time delay upon startup after power failure. Default is 0.
Temperature Offset: This is used for calibration. 
Range is –10°C to +10°C. Default is 0.
Backup System Settings: This screen allows the user
to select the backup type and backup setpoint for units
with a backup system installed. For more information,
refer Backup System (Optional).
Press the Temperature Setpoint button to display the
Setpoints screen:
Figure 36. Setpoints Screen
20 |
Operation
The setpoint and temperature alarm parameters may be
adjusted by swiping the spin control up / down or
pressing the up / down arrows.
Note: A setpoint change may automatically change the warm
and / or cold alarm setpoints as well to prevent unnecessary
alarms.
After selecting the temperatures, press the Save button
to confirm changes.
Note: If the save button is not pressed, the unit will not
respond to the setpoint change request.
Warm Alarm Test: Clicking this button puts the system into a
warm alarm test which simulates a warm alarm experience.
Once this is selected, the home screen is displayed with the
current temperature readout. The temperature display will
increase to the warm alarm temperature setpoint. Once the
warm alarm temperature is reached, the user is prompted to
end the test.
Connectivity
The Connectivity tab is used to setup a wireless network and
connect to InstrumentConnect. Ensure to follow both steps to
store information on InstrumentConnect.
The following screen is displayed when the Connectivity
button is selected.
Figure 37. Connectivity Screen
Ultra Low Temperature Freezers

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents