Thermo Scientific TSX Series Installation And Operation Manual page 56

Ultra low temperature freezers
Table of Contents

Advertisement

Alarm Summary
Alarm Message
Refrigeration
System Failure
Inactive
Wrong Power Alarm
Wrong Power Alarm
Inactive
Wrong Model Alarm
Wrong Model Alarm
Inactive
Firmware Build
Incompatible
Ultra Low Temperature Freezers
Remote
Heart
Tone
Alarm
Status
Event
None
Red
None
High
Red
Yes
None
Yellow
None
High
Red
Yes
None
Yellow
None
High
Red
Yes
*
Ringback
Message Details
Message Header: System Refrigeration
Failure Recovered
Message Details: An error in refrigeration
None
system has been recovered.Select check box
and press Acknowledge to clear this
notification.
Ticker Message: None
Message Header: Wrong Power Alarm
Message Details: The unit has detected the
None,
wrong power connected. Verify the proper
constant
voltage.
audible
Ticker Message: Wrong Power Alarm. Press
bell icon for more information.
Message Header: Wrong Power Recovered
Message Details: Wrong Power has been 
None
recovered. Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
Message Header: Invalid Control Model
Alarm
Message Details: Invalid Control Model
None,
Alarm. Contact service to ensure the correct
constant
model is selected for the system to avoid
audible
cargo loss.
Ticker Message: Invalid Control Model
Alarm. Press bell icon for more information.
Message Header: Invalid Control Model 
Recovered
Message Details: Invalid Control Model has
None
been recovered.Select check box and press 
Acknowledge to clear this notification.
Ticker Message: None
Message Header: Firmware Build
Incompatible Alarm
Message Details: Firmware build indicates 
incompatibility that can result in modules to
Yes
be non-coherent.
Ticker Message: Firmware build
incompatible. Press bell icon for more
information.
Appendix A: Alarm Summary | 51