Thermo Scientific TSX Series Installation And Operation Manual page 55

Ultra low temperature freezers
Table of Contents

Advertisement

Alarm Summary
Alarm Message
Compressor 
Temperature
Inefficient Buck /
Boost
Power Failure Alarm
Power Failure 
Inactive
Clean Filter Active
Refrigeration
System Failure
50 |
Appendix A: Alarm Summary
Remote
Heart
Tone
Alarm
Status
Event
None
Yellow
None
None
Yellow
None
High
Red
Yes
None
Yellow
None
None
Yellow
None
High
Red
Yes
*
Ringback
Message Details
Message Header: Compressor Temperature
Notification
Message Details: A refrigeration system 
temperature has exceeded the compressor 
None
temperature setpoint of XX.X C. Contact 
customer service.
Ticker Message: None
Message Header: Inefficient buck / boost 
notification
Message Details: Freezer input voltage is
None
out of range. Ensure proper supply voltage is
applied. If unable to clear this notification,
contact customer service.
Ticker Message: None
Message Header: Power Failure Alarm
Message Details: Unit in power failure mode. 
Display operating on battery power. Check
Yes
unit plug, unit circuit breaker in the ON
position, and supply voltage.
Ticker Message: Power Failure Alarm Active.
Press bell icon for more information.
Message Header: Power Failure Recovered
Message Details: Power has been restored. 
None
Press the yellow triangle to clear this
notification.
Ticker Message: None
Message Header: Clean Filter Notification
Message Details: Recommend to clean the
None
filter and condenser. If unable to clear
notification, contact customer service.
Ticker Message: None
Message Header: System Refrigeration
Failure Alarm
Message Details: An error has occurred
within the refrigeration system. Contact
Yes
customer service.
Ticker Message: System Refrigeration
Failure Alarm. Press bell icon for more
information
Ultra Low Temperature Freezers

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents