Alarm Summary
Alarm Message
TC1-9 has been
Main to UI Lost
Communication
Main to UI Lost
Communication
Inactive
BUS Lost
Communication
BUS Lost
Communication
Inactive
Failure to Reach
Setpoint
Ultra Low Temperature Freezers
Remote
Heart
Tone
Alarm
Status
Event
None
Yellow
None
High
Red
Yes
None
Yellow
None
High
Red
Yes
None
Yellow
None
None
Yellow
None
*
Ringback
Message Details
Message Header: TCXX Out of Range
Notification Recovered
Message Details: Information TCXX has
None
recovered. Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
Message Header: Lost Communication
Alarm
Message Details: A communication error
Yes
has occurred within the system. Contact
customer service.
Ticker Message: Lost Communication Alarm
Active. Press bell icon for more information.
Message Header: Lost Communication
Alarm Recovered
Message Details: A communication error
None
has recovered. Select check box and press
Acknowledge to clear this notification.
Ticker Message: None
Message Header: Backup System
Communication Alarm
Message Details: A communication error
has occurred within the back up system.
Yes
Contact customer service.
Ticker Message: BUS Lost Communication
Alarm Active. Press bell icon for more
information.
Message Header: BUS Lost Communication
Alarm Recovered
Message Details: A BUS communication
None
error has recovered. Select check box and
press Acknowledge to clear this notification.
Ticker Message: None
Message Header: Unable to Reach Setpoint
Notification
Message Details: Door openings or product
None
loading may cause this notification. Allow unit
to stabilize. If condition persists, contact
customer service.
Ticker Message: None
Appendix A: Alarm Summary | 49