Alarm Summary
Alarm Message
Control Probe
Failure
Control Probe
Failure has been
Heat Exchange
Probe Failure
Heat Exchange
Probe Failure has
been
TC1-9 Failure
48 |
Appendix A: Alarm Summary
Remote
Heart
Tone
Alarm
Status
Event
High
Red
Yes
None
Yellow
None
High
Red
Yes
None
Yellow
None
None
Yellow
None
*
Ringback
Message Details
Message Header: Control Probe Failure
Alarm
Message Details: Cannot display cabinet
temperature. The freezer will continue to
Yes
operate in full run mode. Contact customer
service.
Ticker Message: Control Probe Failure
Alarm active. Press bell icon for more
information.
Message Header: Control Probe Failure
Recovered
Message Details: Control Probe Failure
None
Alarm has been cleared. Press the yellow
triangle to clear this notification.
Ticker Message: None
Message Header: Heat Exchanger Probe
Failure Notification
Message Details: The freezer will continue to
operate with current freezer set points, but
cabinet temperature variation will increase.
Yes
Contact
customer service.
Ticker Message: Control Probe Failure
Alarm active. Press bell icon for more
information.
Message Header: Heat Exchanger Probe
Failure Recovered
Message Details: Heat Exchange Probe
None
Failure Alarm has been cleared. Press the
yellow triangle to clear this notification.
Ticker Message: None
Message Header: TCXX Out of Range
Notification
Message Details: Information TC's has
malfunctioned. This doesn't affect the
None
performance of the unit. Contact service for
further assistance.
Ticker Message: None
Ultra Low Temperature Freezers