6 Commissioning
6.1 Commissioning the device
Prerequisites
● System components correctly installed
● Vacuum- or system-related conditions satisfied
● Technical data observed
● Mains plug removed
● Required vacuum available
Procedure
1. Connect the wiring for the system components.
2. Establish a LAN connection to the PC (direct or via a network).
3. Set up the PC as specified in the software documentation.
4. Refer to the relevant operating instructions for other system components.
6.2 Switch on the device
Danger to life due to electric voltage
High voltages are present inside the device. When touching parts that are live, there is a risk of
death. If there is visible damage, there is a risk of death when commissioning the device.
► Work on the open device must only be carried out by trained specialist personnel.
► Before carrying out any installation and maintenance work, switch the device off and disconnect
► Never open the device with the current supply connected.
► Secure the current supply against unauthorized or unintentional reactivation.
► Do not insert any objects into the vent openings.
► Never open an external power supply pack.
► Never operate an open or defective device.
► Secure a defective device against accidental operation.
► Protect the device against moisture.
Factory setting
Pfeiffer Vacuum configures complete systems optimally at the factory. Therefore, do not
change anything without good reason.
─ < 10
-4
hPa for QMA 250 F
─ < 10
-5
hPa for QMA 250 M
DANGER
it from the current supply.
– After switching off, wait about 60 seconds and then disconnect all cables (power cable at the
end).
PrismaPro Web UI does not allow data storage
PrismaPro Web UI is a remote monitoring and diagnostic tool, and is designed for data
monitoring and simple vacuum diagnostics. The storage of data or display of data in a trend
graph is not possible with PrismaPro Web UI.
Use a complete software program such as PV MassSpec for the storage of data, the se-
lected peak mode or trend graphs.
Only switch on or switch off the QME at the mains plug
Switch the QME on or off by inserting or removing the mains plug only. Always leave the
24 V connector plugged in.

Commissioning

39/86