6. Troubleshooting

6.2 Error Messages

Message
F RAM, SEEP ROM
Check!
Over Temprature!
Doorlock Error
Thermal Sensor Error
Hi Limit Temperature!
Low Limit
Temperature!
Start RPM Error
Low RPM Error
Over RPM Error
Invert Fail
Imbalance Error
Technical Support :+82-2-3452-8966 / techsupport@ihanil.com
Cause
Occurs the data storage memory failure.
Contact Technical Support.
Occurs the temperature is above the allowable limit.
Contact Technical Support.
Occurs the lid sensor failure.
Contact Technical Support.
Occurs the temperature sensor failure.
Contact Technical Support.
Occurs the current temperature is higher than temperature
limit range.
Contact Technical Support.
Occurs the current temperature is lower than temperature
limit range.
Contact Technical Support.
Occurs when 100 RPM is not reached within 2 seconds after
start.
Contact Technical Support.
Occurs when the setting rpm is more than + 2%.
Contact Technical Support.
Occurs when the setting rpm is more than - 2%.
Contact Technical Support.
Occurs when an inverter failure.
Contact Technical Support.
Occurs wheh imbalnce is detected or imbalance sensor
failure.
- Make sure the loaded blood bag adaptors opposite each
other on the rotor are the same weight/volume
- Contact Technical Support.