Error Messages -15; Emergency Lid Open -16 - Hanil Component R6 Operating Manual

Large capacity centrifuge
Hide thumbs Also See for Component R6:
Table of Contents

Advertisement

6. Troubleshooting
6.2 Error messages
Error
No display
No running
The lid is not
open
The lid is not
closed
On
acceleration,
the device is
vibrating
Failure of
power supply
Etc
15 l www.ihanil.com
Display
Cause
- No power connection
None
- Blackout
- No rotor
Unknow
- Errors in drive or rotor
Rotor ID!
recognition function
Hi
-The current temperature
Temperature
is higher than the set
Error!
temperature
Low
-The current temperature
Temperature
is lower than the set
Error!
temperature.
None
-Blackout
To close the
- No working of lid pin
door
- No power connection
-Rotor is loaded
** Imbalance
asymmetrically
Error! **
- Not tighten the rotor
with shaft
** Power
- Failure of power supply
Failure! **
** Motor
- Motor overheat
Overheat! **
** Over
-The current speed is
Speed! **
higher than the set speed
Solution
- Check the power connection
- Check the main fuse in
laboratory and device
- Load a rotor
- Restrat the device
-Check the rotor sensor for
foreign substance.
- Set the temperature higher
- Set the temperature lower
- Emergency lid open
- Check the power connection
- Close the lid
- Load the rotor symmetrically
- Tighten the rotor to the
motor shaft.
- check the outlet
- Stop the centrifugation
- Let the device cool down
- Stop the centrifugation
- Repeat the run