Siemens SITRANS LR250 Operating Instructions Manual page 120

Radar transmitters
Hide thumbs Also See for SITRANS LR250:
Table of Contents

Advertisement

2.6.3. Channel
Used to select between the different Level Transducer Block outputs.
Options
LEVEL/VOLUME
LEVEL
DISTANCE
2.6.4. Input Scaling
2.6.4.1.
Defines the operational lower range value of the input value (Process Value
Scale) in PV (volume/level) Units. Process Value Scale normalizes the input
value to a customer-defined range.
Values
2.6.4.2.
Defines the operational upper range value of the input value (Process
Value Scale) in PV (volume/level) Units. Process Value Scale normalizes
the input value to a customer-defined range.
Values
1)
The point from which High and Low Calibration points are referenced: see Dimensions
on page 11 and Threaded Horn Antenna with extension on page 12.
Page 114
SITRANS LR250 (FF) – OPERATING INSTRUCTIONS
Description
Level value converted to Volume
[through Linearization (2.4.)]
*
Level value
Distance value
distance
level
Lower Value
Upper Value
100
80
60
40
20
0
1
2
3
Output Scale
high calibration
point (process
full level)
Range: -999999 to 999999
Default: 0 %
Range: -999999 to 999999
Default: 100 %
Provides Output values
(Out) to AIFB 1 or AIFB 2
4
5
6
1)
Reference point
Low Calibration Point
Low Calibration Point
Sensor Reference Point
1)
sensor reference point
low calibration point
(process empty level)
7ML19985KL01