Replacing The Fuse; Remote Maintenance - Palas Fidas 200 Series Operating Manual

Fine dust measuring device
Table of Contents

Advertisement

7  Maintenance
Note
98
7.6

Replacing the fuse

Danger to life due to electric shock
Mains voltage is applied to the electrical connections of the device.
4 Switch off the power supply
4 Secure the power supply against accidental reconnection
Two fuses (fine wire fuse M2A) are built into the device to protect it against
overvoltage. The fuses can be accessed on the rear of the device. The fuses
are located in the fuse holder in the socket of the power plug.
To replace the fuses, proceed as follows:
1.
Shut down the firmware and switch off the device.
2.
Disconnect the power plug.
3.
Pull out the fuse holder using a small screwdriver.
4.
Replace the fuse.
5.
Reinsert the fuse holder.
6.
Reinsert the power plug.
Check whether the device starts up as usual and operates normally. Pay at-
tention to the startup of the device in order to rule out any damage to the
components – especially if the fuse was blown. If any irregularities are de-
tected, contact the manufacturer.
7.7

Remote maintenance

We recommend registering on the Palas website to obtain access to the
password-protected area.
4 Log in with your user name and the corresponding password.
This will grant access to software and firmware updates, remote device
maintenance and creation of a support ticket.
DANGER
5200-en_V2.0_06/20