Palas Fidas 200 Series Operating Manual page 37

Fine dust measuring device
Table of Contents

Advertisement

5200-en_V2.0_06/20
Attaching and connecting sampling tube
There is a risk of pinching and shearing points when placing the
sampling tube on the aerosol inlet tube.
Risk of injury to unprotected fingers.
4 Wear protective gloves.
1.
Hold the sampling tube up to light to verify that there are no foreign ob-
jects in the tube.
2.
Carefully guide the sampling tube over the aerosol inlet tube on the
controller.
Important: There should be a gap of 1 to 2 mm between the sampling
tube and the contact surface. If the tube rests directly on the contact
surface, it can subject the aerosol sensor to mechanical strain.
3.
Secure the sampling tube with the pipe clamp. Ensure that the tube is
positioned vertically.
4.
Screw in the union nut and tighten by hand.
5.
Plug the cable from the drying section IADS (on the sampling tube) into
the receptacle "IADS" on the back of the controller.
Attach the sampling head.
For more information, refer to "Attaching sampling head [} 38]."
4  Initial commissioning
CAUTION
37