Palas Fidas 200 Series Operating Manual page 69

Fine dust measuring device
Table of Contents

Advertisement

Parameter
dusttype=
sensor_selection=
Automated_cleaning=
Alarm_threshold=
Alarm_value=
Alarm_email_address=
Password_service
Calibration_IADS_re-
strict=
Calibration_tempera-
ture=
Channel_deviation=
Weatherstation_con-
nected=
Weatherstation_com-
port=
Weathersta-
tion_scale_T=
Weathersta-
tion_scale_p=
Weathersta-
tion_scale_h=
Weatherstation_off-
set_T=
Weatherstation_off-
set_p=
Weatherstation_off-
set_h=
Weatherstation_equa-
tion=
Discmini_connected=
Discmini_comport=
Discmini_interval=
Sensor_selection=
PM_interval=
IP_UDP_broadcast=
5200-en_V2.0_06/20
Entry
Meaning
2
Default. Please do not change.
2
Default. Please do not change.
yes / no
Yes = automatic cleaning is activated.
3
99999 µg/m
Limit of PM fraction that controls the digital alarm (digital
out)
PM10
Specification of PM fraction that controls the digital
alarm (digital out)
""
E-mail address to which a message is sent in the event
of an alarm.
-1
Password for "Expert user mode"
yes / no
Yes = calibration is not enabled until the setpoint is
reached.
50
Setpoint for drying section IADS at which calibration is
enabled.
3.5
Default. Please do not change.
yes / no
Yes = the weather station is connected.
4
COM port of connected weather station.
1
Entry of possible slope scales for adjusting the weather
station; obtained from comparison to transfer standard.
1
1
1
Entry of possible offset factors for adjusting the weather
station; obtained from comparison to transfer standard.
1
1
X_corr = scale *
Function applied to calibrate the weather station.
x + offset
yes / no
Yes = discmini is connected.
81
COM port for discmini.
300 s
Interval of reported discmini files.
2
Default. Please do not change.
900 s
Interval for sliding average for PM fractions. 900 sec-
onds pursuant to certificate.
127.0.0.1
UDP address for data transmission.
5  Operation and settings
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fidas 200Fidas 200 sFidas 200 e

Table of Contents