5.2.15 Device Status Screen - Palas Fidas 200 Series Operating Manual

Fine dust measuring device
Table of Contents

Advertisement

5  Operation and settings
Menu path
Parameter
sensor flow
coincidence
58

5.2.15 Device status screen

The device status screen shows the status of device components. The
status information is saved with every data set. Observe the device status to
be able to detect tendencies and properly plan maintenance.
Depending on the communication profile used, status data may be output via
the interface. This is the case e.g. with the communication profile "Bayern-
Hessen."
If a value exceeds the specified limit, the device changes to "check de-
vice status" mode. In this case the source of the problem has to be
found and remedied. The table below contains various explanations.
Main menu > device status
Device status screen
Meaning
Explanation
The measured
flow in between indicates the current flow rate. The set-
data is invalid.
point is 4.8 l/min. The flow rate is standardized to standard
atmospheric temperature and pressure (SATP), i.e. based on
25°C and 1013 hPa.
If the actual value deviates more than 5% from the setpoint,
the measured data is invalid. An error indication appears.
flow velocity in between shows the detection volume,
which is the amount of air that is tested.
An error is indicated if the particle velocity deviates more than
15% from the setpoint.
The setpoint is the value velocity_calibrated entered in
the promo.ini file.
The measured
Detection of more than one particle in the optical detection
data is invalid.
volume.