Palas Fidas 200 Series Operating Manual page 85

Fine dust measuring device
Table of Contents

Advertisement

5200-en_V2.0_06/20
This is how to use the calibration dust properly:
1.
Shake the closed container gently.
ð Some of the particles swirl in the air inside the container and remain
in the air for a few minutes.
2.
Unscrew the lid and place the end of the adapter hose in the container.
The hose should protrude only a few centimeters into the container,
and it should not touch the sides or bottom of the container.
This is the only way to ensure that only a few dust particles are drawn
into the hose.
Proper handling of MonoDust 1500
6  Verifying and calibrating measuring system
85