5 Operation and settings
Menu path
52
5.2.11 Alarms screen
Automatic notifications can be set up in this screen.
You can choose to receive an e-mail message or set an alarm on the digital
output of the device "digital_out."
Main menu > accessories > alarms
Alarms screen
Notification by e-mail
If this feature is activated (X in the box), the device sends a message to the
stated address if one of the status parameters exceeds the limits. Refer to
the section "Device status screen [} 58]."
Alarm signal at digital output
If this feature is activated (X in the box), the device sends a signal to the dig-
ital output if the limit specified in the file promo.ini is exceeded:
Parameters for the limit: alarm_threshold=
Parameters for the OM fraction: alarm_value=
Default setting for alarm_threshold= 99999
Default setting for alarm_value = PM10
user@company.com
5200-en_V2.0_06/20