Checking Device For Leakage - Palas Fidas 200 Series Operating Manual

Fine dust measuring device
Table of Contents

Advertisement

6  Verifying and calibrating measuring system
78
6.3

Checking device for leakage

Net leakage of total system
The net leakage is the difference in the parameter sensor flow rate
when the pumps are switched with the IADS tube closed (= "gross leakage")
and when the pumps are switched off (= "offset").
Net leakage = gross leakage – offset
When the net leakage is less than 0.5 l/min, the entire device is considered
to be sealed.
If leakage is greater, check the following:
• Seat of IADS tube in aerosol inlet
• Seals on aerosol inlet
When checking, the IADS tube has to briefly be sealed with a plug.
Prerequisites for testing device leakage
• Calibration mode is activated.
• The controller has been tested for leakage and verified.
• The screen Main menu > expert user menu is visible.
5200-en_V2.0_06/20