3. Select to perform spot or sweep correction through ▲ or
Figure4-3.
"SPOT" means only open correction for the current frequency point; while"SWEEP" means
open correction for all frequency points.
)Note Press
4. Press ESC key to cancel correction and exit to test status.
5. Press ENTER key to start correction.

4.2.9 Short correction

Press SHIFT + SHORT to enter short correction menu.
AD2817C's open correction can eliminate the effect of test cable, test fixture and stray
impedance in series with the device under test, such as lead resistance or lead inductance.
Perform the following steps to perform short correction:
1. Press SHIFT key at test status, and "SHIFT" at left corner is on.
2. Press SHORT key to select correction function, as Figure 4-2 shows on LCD.
3. Select to perform spot, sweep or DCR correction through ▲ or ▼
in "SPOT" means only short correction for the current frequency point; while"SWEEP"
means short correction for all frequency points; "DCR" means DC resistance short
correction.
)Note Press
4. Press ESC key to cancel correction and exit to test status.
Figure 4-2 Open spot correction
Figure 4-3 Open sweep correction
to enter short correction menu.
Figure 4-4 Short spot correction
to enter open correction menu.