Leica SPIDER GR30 User Manual page 52

Hide thumbs Also See for SPIDER GR30:
Table of Contents

Advertisement

GR50 setup in a non-
DHCP network using
display and buttons
52
Step
Description
8.
Press OK.
9.
Open a browser window and enter 192.168.0.3 to open the web
interface.
Per default, the GR30/GM30/GR50 instrument is configured to obtain
an IP address automatically from a DHCP network. To use the default
static IP address 192.168.0.3, reboot the GR30/GM30/GR50 once it
is connected to the crossed Ethernet cable.
In a field campaign the GR50 can be configured through the display and but-
tons, to start or stop pre-configured data streams and logging sessions.
For initial setup, the GR50 IT configuration for setup in a non-DHCP network
can be done through the display and buttons.
002981_003
Step
Description
1.
Turn on the GR50.
2.
Use the arrow buttons, go to Configuration, Site Config. Enter the
coordinates, antenna type and site code.
3.
Press Enter to store all changes.
4.
Use the arrow buttons go to Configuration, Network Config. Enter
the IP address, Subnet mask and gateway of the network to connect
the instrument to.
5.
Press Enter to store all changes.
a
Left button
a
b
Right button
b
c
c
Up button
d
d
Cancel button
e
e
Down button
f
f
Enter button
g
g
Display
h
h
ON/OFF button