Using Your Kane458S For The First Time; Using Your Kane458S Every Time; Fresh Air Ourge On Switch On - Kane 458s Manual

Flue gas analyser with direct co2 measurement and co sensor protection
Table of Contents

Advertisement

USING YOUR KANE458s FOR THE FIRST TIME

Charge your analyser's batteries for 8 hours – an overnight charge
should be sufficient for an average 8-hour day.
Take time to read this manual fully & be aware your analyser's
configuration may not support all features detailed in this manual.
Take time to set up your requirements before using your analyser.

USING YOUR KANE458s EVERY TIME:

Check your analyser's water trap is empty & particle filter is not dirty.
To empty, pull out the red plug, shake out water & replace red plug.
To change the particle filter, remove protective rubber cover, slide the
water trap from analyser, remove particle filter from its spigot &
replace.
Reconnect your analyser's water trap & rubber protective cover.
Connect your flue probe hose to your analyser's flue gas inlet – see p.8.
Connect your flue probes temperature plug to your analyser's T1 socket
– check plug orientation is correct – See p.8.
FRESH AIR PURGE ON SWITCH ON
Your analyser starts a zero countdown to auto calibrate itself.
When finished, rotate dial to "RATIO" – In fresh air CO reading should be
zero.
Rotate dial to "O2/EFF" – In fresh air O2 readings should be 20.9% +
0.3%.
12 KANE 458s MANUAL