MEASURING FLUE GASSES

After the countdown is finished and the analyser is correctly set up, put
its flue probe into the appliance's sampling point. The tip of the probe
should be at the centre of the flue. Use the flue probe's depth stop cone
to set the position.
With balanced flues, make sure the probe is positioned far enough into
the flue so no air can "back flush" into the probe.
WARNING
Ensure the flue probe handle does not get hot!
Make sure you do not exceed the analyser's operating specifications. In
particular:
• Do not exceed the flue probe's maximum temperature (600°C)
• Do not exceed the analyser's internal temperature operating range
16 KANE 458s MANUAL