RaySafe Xi User Manual – Survey Measurement

SURVEY MEASUREMENT

The RaySafe Xi survey detector is designed for measuring leakage or scattered radiation from X-ray equipment
and γ-emitting isotopes.
Available dose quantities (selectable from SETUP MENU) are:
• Air kerma (K
), in Gy
air
• Exposure (X), in R
• Ambient dose equivalent (H*(10)), in Sv
1. SENSOR MENU
R/F
For leakage or scatter from R/F equipment.
Mammography
For leakage or scatter from mammography equipment.
When measuring on γ-emitting isotopes.
Nuclear Med.
After an offset adjustment, the detector waits in IDLE MODE. Start measuring by a short press on the RaySafe
Xi survey detector key.
2. MEASURE MODE
In MEASURE MODE the ticker frequency is proportional to current dose rate. Measured values are continuously
updated on the display (twice per second).
26