RaySafe Xi User Manual – mA/mAs Measurement
mAs
mAs
mA
mA
Time
ms, s
Pulses
pulses
frames/s
Frame rate
mAs per frame
mAs/f
3. SETUP MENU
Press SELECT to enter the SETUP MENU (from MEASURE MODE) and STEP to step between setup
parameters.
When in the SETUP menu, the active selection will blink. All values are stored in a non-volatile memory and are
valid until manually changed. At start up (after the battery status information), valid Trig delay and Calc. delay
are displayed in sequence. (Factory settings in bold.)
Trig delay
A delay in ms after the normal trig of the RaySafe Xi. Utilize when an unwanted part
of an exposure, such as a pre-pulse, should be excluded from the measurement. No
measurements are performed during the Trig delay. (0, 5, 10, 50, 100, 200, 500, 1000,
2000 ms)
Calc. delay
Defined as the dead time after end trig, before data is calculated. Default is 0.5 s but it is
recommended to use a longer delay when measuring on pulsed fluoroscopy where the
time between pulses may exceed 0.5 s. (0.5, 2, 4, 6 s)
Display lighting option. (off, 2, 5, 10 s, on)
Backlight
Auto Power Off
Never, or after 5, 20 or 60 minutes of inactivity.
Info
Displays the serial number (S/N) and firmware version of the detector and the RaySafe Xi
base unit and the calibration date for the selected sensor. There is also a possibility to
reset the instrument settings to factory defaults.
30