TRANSFER DATA TO A COMPUTER

To view the waveform in the RaySafe Xi view and/or transfer measured data to a PC, connect a serial cable or
Bluetooth module (option) to the communication port on the RaySafe Xi base unit.

BATTERY AND CHARGER

CAUTION!
Only authorized personnel may remove and replace the battery.
Fully charge the battery with the provided battery charger before first use. Battery status in percentage is
displayed, together with setup information, when the instrument is switched on.
RaySafe Xi/Unfors Xi Platinum Plus
FOR RAYSAFE Xi
The provided battery charger may be connected during all measurements.
FOR UNFORS Xi PLATINUM
The provided battery charger may be connected during measurement except when measuring mAs, as it may
create ground currents affecting the measurement. However, the RaySafe Xi battery charger may be connected
during all measurements.
Xi version
RaySafe Xi
Unfors Xi Platinum
Unfors Xi Platinum
Battery type
Hours of operation
7.4 V Li-ion
20–40 hours
9 V NiMH
10–20 hours
Charge time
4 hours
15 hours
9