3. SETUP MENU
Press SELECT to enter the SETUP MENU (from MEASURE MODE) and STEP to step between setup
parameters.
All values are stored in a non-volatile memory and are valid until manually changed. At start up (after the
battery status information), valid Calc. delay is displayed.
(Factory settings in bold.)
Calc. delay
Defined as the dead time after end trig, before data is calculated. Default is 0.5 s but it is
recommended to use a longer delay when measuring on pulsed fluoroscopy where the
time between pulses may exceed 0.5 s. (0.5, 2, 4, 6, 7 s)
Dose unit
Change the default units and/or use this feature to recalculate a value to a different unit.
(Gy or R per s, m or h, where 1 Gy = 114.1 R.)
Backlight
Display lighting option. (off, 2, 5, 10 s, on)
Auto Power Off
Never, or after 5, 20 or 60 minutes of inactivity.
Info
Displays the serial number (S/N) and firmware version of the connected detector and
the RaySafe Xi base unit and the calibration date for the selected sensor. There is also a
possibility to reset the instrument settings to factory defaults.