Reset Switch; Vertical Detector Placement; Kvp On Amx; Display Codes - Fluke Biomedical RaySafe Xi 8201011 User Manual

Table of Contents

Advertisement

RaySafe Xi User Manual – Tips and Tricks

RESET SWITCH

If the software hangs up, by any reason, activate the system reset by pressing a thin object through the small
hole beside the RS-232 connector on the side of the RaySafe Xi base unit.
If the detector hangs up when the battery is down and does not respond even when a charger is connected,
reset as described above and charge for some minutes before turning the instrument on again.

VERTICAL DETECTOR PLACEMENT

When measuring on a mammography machine with a horizontal beam, such as MultiCare Platinum, let the
pressure from the compression paddle fix the RaySafe Xi mammography detector in position. Put the detector
with the cable end pointing downwards and with the MAM-window centered in the compression paddle biopsy
window, as shown below.

kVp ON AMX

When measuring on AMX machines with R/F high, keep a short distance to get kVp instead of kV.

DISPLAY CODES

Display
* * *
- - -
< 1 ms
Meaning
Measurement has started.
No value can be calculated.
Less than 1 ms valid samples.
43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents