RaySafe Xi User Manual – Mammography Measurement

MAMMOGRAPHY MEASUREMENT

6 cm
Place the detector along the X-ray tube axis (with the cable end pointing towards the X-ray machine). For kV
measurements, make sure that the center of the MAM sensor square is placed 6 cm from the examination table
front edge. Dose and HVL can be measured from other positions as well.
The following concerns Unfors Xi Platinum, not RaySafe Xi:
NOTE!
When measuring on mammography machines with carbon fibre table, we recommend not to
use the power adapter to charge Xi, or use something that creates at least a 5 mm distance
between the table and the detector, such as a plastic plate.

SPECIAL MEASUREMENT CASES

The table below indicates when to use filter for kV measurements.
Beam quality
Mo/Mo
Mo/Rh
W/Rh
W/Al 20–39 kV
W/Al 40–49kV
W/AlScanning
Mo/Rh
When measuring kV with Mo/Rh, use the included 2 mm Al filter. Follow the instructions printed on the filter.
Filter
No filter
2 mm Al
No filter
No filter
2 mm Al
2 mm Al
Comment
Use the R/F high detector
14