RaySafe Xi User Manual – mA/mAs Measurement
NOTE!
Use a laptop operating on battery or a Bluetooth connection when capturing waveforms or
transferring data to a computer running RaySafe Xi view. Connection of a PC (mains supplied)
via a serial cable to the RaySafe Xi base unit may create ground currents affecting the
measurement.
For Unfors Xi Platinum: Also avoid connecting the Unfors Xi battery charger when measuring
mA/mAs, as it may create ground currents affecting the measurement. However, the RaySafe
Xi battery charger may be connected during all measurements.
Be aware of rush currents (loading the capacitance in the high voltage cables) in the initial part of the exposure.
Use a Trig delay of 5 ms in the SETUP MENU to exclude the rush current from the measurement, if desired.