4.3

Grounding

Ensure continuity (Resistance less than 0.2 MΩ) between the grounding
terminal and grounding pole in Figure 4-6 or Figure 4-7. Grounding resistance
100 Ω or less is recommended. If continuity between the ground terminal and
the ground electrode cannot be ensured, use the following cables. Do not
share the ground wiring of the XS530 with other devices.
n Applicable Cables
Insulated cables for industrial equipment such as;
• 600V polyvinyl chloride insulated wires (IV): JIS C3307
• Polyvinyl chloride insulated wires for electrical apparatus (KIV): JIS C3316
• 600V grade heat-resistant polyvinyl chloride insulated wires (HIV): JIS
C3317
• Heatproof vinyl insulated wires VW-1 (UL 1015/UL 1007)
Wire size
• Core: AWG 14 to 13 (2 to 2.6 mm
Termination
• Use a round crimp terminal for M4 terminals with an insulation sleeve
Grounding
Terminal
Figure 4-6 Ground Terminal
CAUTION
Grounding is required for safe operation.
)
2
Grounding
Terminal
F10.ai
Figure 4-7 Ground Terminal on bracket side