Page 1
Instruction Model SX42 Manual 2-electrode Conductivity Sensors For High temperature/pressure IM 12D07J03-01EN-P edition...
Page 2
родния ви eзик, ce свържете c най- artículos antiexplosivos en su idioma local, близкия oфис или представителство на deberá ponerse en contacto con la oficina фирма Yokogawa. o el representante de Yokogawa más cercano. (CZ) Všechny uživatelské příručky pro výrobky, na něž se vztahuje nevýbušné schváleni Tous les manuels d’instruction des...
Page 3
Ex, deverá entrar em contacto com a delegação mais próxima (SLO) ou com um representante da Yokogawa. Vsi predpisi in navodila za AEX Ex sorodni pridelki so pri roki v anglišèini. Èe so Ex sorodna navodila potrebna v vašem tukejnjem jeziku, kontaktirajte vaš...
Amphenol connector to fit that were ordered are included and correct. the Yokogawa WU40 cable, or with a Variopin connector to fit with Yokogawa For some specific sensor information, the WU10/WE10-cable.
The following information must be 1.4 Serial number included in the letter accompanying the The Serial number is defined by nine (9) returned goods: alphanumeric characters: • Model Code and Serial Number. Production location • Original Purchase Order and Date. Year/Month code •...
2.5 Functional specifications (at 25 °C) Temperature element : Pt1000 to IEC 751 Nominal Cell Constant : SX42-SX24 = 0.1 cm SX42-SX34 = 0.01 cm Note 1 : The temperature sensor included in the sensor is designed for process compensation and for indication. It is NOT designed for process temperature control.
2.7 Operating range Conductivity range* at actual process temperature : 1 μS x C.C. – 200 mS x C.C. See Figure 1 * measurement range dependent on input range analyzer. Figure 1: Measuring range of conductivity sensors section Temperature @ 1 Bar (14.5 PSIG) : Threaded models (-BS, NS) : 0 °C to 200 °C (32 °F to 392 °F) Threaded models (-BV, NV)
Figure 2: Pressure vs Temperature Cable length: Threaded models (-BS,-NS) : max. 60 meter with WU40 cable in combination with WF10 cable and BA10 junction box Threaded models (-BV,-NV) - Variopin connector connected to : max. 60 meter with WU10-V-D/WE10 (possibly FLXA analyzer.
Note 2 : Sensor input circuits when connected to an analog intrinsically safe certified SC transmitter (e.g. Yokogawa transmitter Model FLX21/FLX202 series or Yokogawa transmitter Model SC202S series): In type of protection intrinsic safety Ex ia IIC maximum input values are: Ui= 14.4 V, Ii= 116.5 mA, Pi= 0.3424 W, Li= 0 mH, Ci= 0 nF (-SX..
Page 13
Note 3 : Ambient temperature: -30°C to +40°C for temperature class T6, -30°C to +95°C for temperature class T5, -30°C to +125°C for temperature class T4 (BV and NV type), -30°C to +130°C for temperature class T4 (BS, NS and *F types), -30°C to +125°C for temperature class T3 (BV and NV type), -30°C to +165°C for temperature class T3 (BS, NS and *F types), -30°C to +125°C for temperature class T2 (BV and NV type),...
Page 14
Specific conditions : See Control Drawing D&E 2020-024-A50. (Page 15-18) of use Temperature classes for SX42 models are defined T2…T6, see Note 5 Note 4 : Intrinsically safe, entity, for Class I, Division 1, Groups A, B, C and D;...
Page 15
Control drawing: D&E 2020-024-A50 (part 1) IM 12D07J03-01EN-P...
Page 16
500 V rms. Sensor Model code: Table 3: Model Suffix Codes Option Codes SX42 -abcd-efgh abcd Cell Constant: Four alphanumeric character (A to Z, 0 to 9 or a hyphen) BS ISO7/1-R1 screw thread + connector without ID-chip...
Page 17
Control drawing: D&E 2020-024-A50 (part 2) IM 12D07J03-01EN-P...
Page 18
Model SA11-C1 Smart Adapter however provides this required isolation. Sensor Model code: Table 4 Model Suffix Codes Option Codes SX42 -abcd-efgh abcd Cell Constant: Four alphanumeric character (A to Z, 0 to 9 or a hyphen) Connection type:...
Page 19
FM-Canada Applying standards : CAN/CSA-C22.2 No. 60079-0 CAN/CSA-C22.2 No. 60079-11 Certificate no.* : FM20CA0062X IS CL I, DIV 1, GP ABCD, T2…T6 CL I, ZN 0, Ex ia IIC, T2…T6 Ga Control Drawing: D&E 2020-024-A51 Electrical data : See Note 6 Specific conditions : See Control Drawing D&E 2020-024-A51.
Page 20
Control drawing: D&E 2020-024-A51 (part 1) IM 12D07J03-01EN-P...
Page 21
500 V rms. Sensor Model code: Table 5 Model Suffix Codes Option Codes SX42 -abcd-efgh abcd Cell Constant: Four alphanumeric character (A to Z, 0 to 9 or a hyphen) Connection type:...
Page 22
Control drawing: D&E 2020-024-A51 (part 2) IM 12D07J03-01EN-P...
Page 23
Model SA11-C1 Smart Adapter however provides this required isolation. Sensor Model code: Table 6: Model Suffix Codes Option Codes SX42 -abcd-efgh abcd Cell Constant: Four alphanumeric character (A to Z, 0 to 9 or a hyphen) Connection type:...
3. INSTALLATION OF SX42 sensors 3.1 Typical installation of SX42 SX42 threaded models are installed by screw-in (ISO 7/1-R 1” or 1” NPT), and the SX42 flanged models by using the pre-mounted flange (DN50 PN63, DN50 PN40 or ANSI 2”...
5. WIRING The SX42 threaded models are provided with a fixed connector. The standard cable used to connect the sensors with Amphenol connector to the analyzer is the WU40. The standard cable used to connect the sensor with VP connector to the analyzer is the WU10/WE10.
Page 27
Note 10 : Preferred connection cable is Yokogawa Model WU10-V-D or Model WE10 The SX42 flanged models are provided with a connection box. The cable used to connect these sensors to the analyzer has to be a high temperature shielded cable. This cable is not supplied by Yokogawa.
Figure 11: Sensor + SA11 Smart adapter Cable mounting When there is less space to install the SA11 on top of the Yokogawa labelled sensor or when the process conditions are higher than +125 °C, an alternative mounting method is to install the SA11 using the optional wall/pipe mounting hardware.
2. Lime, hydroxides or similar sensor. The cell constant value can be applications: 5 % solution of hydrochloride entered directly in the Yokogawa acid. analyzer. The procedure is explained in 3. Organic (e.g. oils, fats) applications: the Instruction Manual of the analyzer .
1-11½ NPT screw thread, plug-socket conn. 1-11½ NPT screw thread, Vario Pin conn. with SENCOM ID-chip DN50-PN63 EN flange DN50-PN40 EN flange 2” 600 LBS ANSI flange Always *A style Style Option 8. SPARE PARTS No spare parts defined for model SX42 IM 12D07J03-01EN-P...
9. CHEMICAL COMPATIBILITY CHART Table12: Note 13 : Information in this list is based on our general experience and literature data and given in good faith. However, Yokogawa is unable to accept responsibility for claims related to this information IM 12D07J03-01EN-P...