Advertisement

User's
Manual
Model ISC40G(S)
Sensors and Fittings for Inductive
Conductivity Measurement
IM 12D08J02-E-E
18th edition

Advertisement

Table of Contents
loading

Summary of Contents for YOKOGAWA ISC40G(S)

  • Page 1 User’s Model ISC40G(S) Manual Sensors and Fittings for Inductive Conductivity Measurement IM 12D08J02-E-E 18th edition...
  • Page 2 свържете c най-близкия oфис или en su idioma local, deberá ponerse en contacto представителство на фирма Yokogawa. con la oficina o el representante de Yokogawa más cercano. (CZ) Všechny uživatelské příručky pro výrobky, na něž...
  • Page 3 Ja vçlaties saòemt Ex ierîèu zgodnych z wymaganiami ATEX, dostępne dokumentâciju citâ valodâ, Jums ir jâsazinâs ar są w języku angielskim. Jeżeli wymagana firmas Jokogava (Yokogawa) tuvâko ofisu vai jest instrukcja obsługi w Państwa lokalnym ję pârstâvi. zyku, prosimy o kontakt z najbliższym biurem Yokogawy.
  • Page 4: Table Of Contents

    Table of Contents 1 INTRODUCTION General Unpacking and Checking Warranty and Service Serial Number definition GENERAL SPECIFICATIONS Measuring elements Materials Functional specifications (at 25°C) Dynamic specifications Operating range Regulatory standards Shipping details Environmental conditions Process connections INSTALLATION OF SENSOR WITH STANDARD OPTIONS 14 Typical installation 3.1.1 Installation of ISC40G(S)-GG (TG) with flange adapters 3.1.2 Installation of ISC40G(S)-TG with T-piece...
  • Page 5 ISC40FF – FLOW FITTING 10.1 General Specifications 10.1.1 Materials 10.1.2 Operating range 10.1.3 Shipping details 10.1.4 Process connections 10.2 Installation of ISC40 sensor in ISC40FF 10.3 Dimensions ISC40FF 10.4 Modelcode ISC40FF 10.5 Spareparts ISC40FF ISC40FD – IMMERSION FITTING 11.1 General Specifications 11.1.1 Materials 11.1.2 Operating range 11.1.3 Process connections...
  • Page 6 IM 12D08J02-E-E...
  • Page 7: Introduction

    (freight 3. Maintenance friendly installation of paid) to the Service Department of the sensors using in-line subassemblies relevant Yokogawa Sales office for repair or 4. Installation in flow chambers for replacement (at Yokogawa’s discretion). measurements in sample streams or...
  • Page 8: Serial Number Definition

    1.4 Serial Number definition The Serial Number is defined by nine (9) alphanumeric characters: X 1 X 2 Production Location X 3 X 4 Year/Month code X 5 X 6 X 7 X 8 X 9 Tracking number Example: N3P600028 Method used for year/month numbering Table 1: Production Year code Table 2: Production Month code...
  • Page 9: General Specifications

    2 GENERAL SPECIFICATIONS The ISC40 inductive conductivity sensor is suitable for use with the Yokogawa inductive conductivity analyzers. 2.1 Measuring elements : Toroïds with high permeability magnetic material Pt1000 or 30k temperature element 2.2 Materials Wetted parts sensor Body ISC40*-G*...
  • Page 10: Operating Range

    Ui = 19.1 V; Ii = 170 mA; Pi = 0.8 W; Ci = 0 nF; Li = 0 mH certified intrinsically safe Yokogawa Inductive Conductivity transmitter model FLXA21 series, model ISC202S series or...
  • Page 11: Shipping Details

    Uo = 19.1 V; Io = 170 mA; Po = 0.8 W; Co≥ Ci + Ccable; Lo≥ Li + Lcable certified intrinsically safe Yokogawa Inductive Conductivity transmitter model FLXA21 series, model ISC202S series or model IC200S series Ambient : T6 for Tamb.
  • Page 12 Control Drawing CSA The ISC40S sensor shall be installed with one of the Yokogawa transmitters model: • ISC202S • IC200S • FLXA21 with following parameters: ISC202S IC200S FLXA21 14.4 V 19.1 V 11.76 V 88 mA 162 mA 60.6 mA ISC40 sensor terminals 11-17.
  • Page 13 Control Drawing FM The ISC40S sensor shall be installed with one of the Yokogawa transmitters model: • ISC202S • IC200S • FLXA21 with following parameters: ISC202S IC200S FLXA21 14.4 V 19.1 V 11.76 V 88 mA 162 mA 60.6 mA...
  • Page 14: Installation Of Sensor With Standard Options

    3 INSTALLATION OF SENSOR WITH STANDARD OPTIONS For optimum measurement results, the ISC40 sensor should be installed in a location that offers an acceptable representation of the process composition and DOES NOT exceed the specifications of the sensor. It is important that the process flow is directed through the hole in the donut of the sensor.
  • Page 15 The sensor cable and mounting thread avoided by using a wrench on the flats are pulled through the hole of the flange, on top of the sensor (see fig. 5). Sensor and the sensor is sealed from the process is installed correctly if the flats are aligned by tightening the mounting nut.
  • Page 16: Installation Of Isc40G(S)-Tg With T-Piece

    3.1.2 Installation of ISC40G(S)-TG with T-piece The ISC40G(S)-TG sensor is a PFA sensor which can be used in combination with a SS304 Nut T-piece of which the innerside is lined with PFA. This can be done using option /TFD (sealing material with stainless steel DN65 DN65 PN10 flange PN10 flange), or by option /TFN (sealing material) if the stainless steel flange DN65...
  • Page 17: Installation Of Isc40G(S)-Gs With Flange Adapters

    3.1.3 Installation of ISC40G(S)-GS plastic nut with flange adapters The ISC40G(S)-GS sensor is designed for tube sanitary applications. For these applications special process connections are necessary. Mounting procedure: • Screw the tube completely in the stainless steel nut. • Thread the sensor cable through ring the flange adapter parts in the right O-ring 18.7 x 2.6...
  • Page 18: Dimensions Isc40 Sensor

    4 DIMENSIONS ISC40 SENSOR Unit: mm (inches) ø17 (0.67”) nut G3/4 For mounting in ISC40PR retractable viton gasket fitting Ø40 (1.57") Minimum hole to fit sensor through: Minimum hole m19 x 1.5 GG Ø48 (1.89") to fit sensor through: Ø48 (1.89") TG Ø50 (1.99”) 55 (2.16") Ø47 (1.85")
  • Page 19: Wiring Isc40 Sensor

    5 WIRING ISC40 SENSOR The ISC40 sensor is provided with a fixed cable. The connections of this dual coax cable and the Yokogawa Inductive Conductivity analyser are given in table 1. Table 1: Definition cable and analyzer Cable wire color...
  • Page 20: Modelcode Isc40 Sensor

    6 MODELCODE ISC40 SENSOR Model Code Suffix Option Description ISC40G General purpose inductive conductivity sensor Sensor Glass filled PEEK, general model type Glass filled PEEK, retractable model Glass filled PEEK, shaft model PFA, general model Temperature Pt1000 sensor 30k thermistor,for IC200 select only T3 Cable length 03 meter 05 meter...
  • Page 21: Spareparts Isc40 Sensor

    7 SPAREPARTS ISC40 SENSOR Parts ISC40 sensor Part no. Description Material Quantity K1500AM Gasket Viton K1500AL Mounting nut AISI 316 SS Options ISC40 sensor, Flange adapters Part no. Description Process connection Material O-ring(s) K1541ZR /SFA 2” ANSI 150 lbs AISI 316 SS Viton K1541ZQ /SFD NW50...
  • Page 22: Eu Declaration Of Conformity

    8 EU DECLARATION OF CONFORMITY YOKOGAWA EU DECLARATION OF CONFORMITY Yokogawa Process Analyzers Europe B.V. Euroweg 2 3825 HD Amersfoort The Netherlands herewith declare under our sole responsibility that the products, model: ISC40G and ISC40S further specified with model suffix- and option codes: As listed in Annex-1 in this document...
  • Page 23: Isc40Fs - Flow Fitting Subassembly

    ISC40FS – FLOW FITTING SUBASSEMBLY General Specifications 9.1.1 Materials Wetted parts Model ISC40FS-FCSA : PVDF (Kynar); Viton Model ISC40FS-PCSA : Polypropylene; Viton Model ISC40FS-SCSA : AISI 316 SS; Viton Model ISC40FS-SCWN : AISI 316 SS; Viton Non-wetted parts : AISI 304 SS 9.1.2 Operating range Temperature Model ISC40FS-FCSA...
  • Page 24: Dimensions Isc40Fs

    O-ring O-ring 40.64 x 5.33 40.64 x 5.33 Holder Holder O-ring O-ring 26.57 x 3.53 26.57 x 3.53 Ring DN50 Ring DN50 Fig 15: ISC40 sensor in screw-in Fig 16: ISC40 sensor in weld-in subassembly ISC40FS-SCSA subassembly ISC40FS-SCWN 9.3 Dimensions ISC40FS Rd 78 x 1/6 Rd 78 x 1/6 Rd 78 x 1/6...
  • Page 25: Modelcode Isc40Fs

    9.4 Modelcode ISC40FS Model Suffix Option Description ISC40FS Flow fitting subassembly Material PVDF Polypropylene Stainless Steel Process Dairy Coupling screw-in* connection Dairy Coupling welded* Thread type NPT or R No thread (for weld-in couplings) Options Material certificate 3.1. EN 10024 (for wetted metal parts only) * Note: according to Din 11851 9.5 Spareparts ISC40FS...
  • Page 26: Isc40Ff - Flow Fitting

    10 ISC40FF – FLOW FITTING 10.1 General Specifications 10.1.1 Materials Wetted parts Model ISC40FF-S : AISI 316 SS; Viton Model ISC40FF-P : Polypropylene; Viton Model ISC40FF-F : PVDF (Kynar); Viton : AISI 304 SS or AISI 316 SS Non-wetted parts 10.1.2 Operating range Temperature Model ISC40FF-S...
  • Page 27: Dimensions Isc40Ff

    70 (2.75) O-Ring 40.64 x 5.33 Holder Ø 10 (0.4) 4x O-Ring 26.57 x 3.53 Ring DN50 pipe mount pipe mount (horizontal) (vertical) Fig 18: Installation of ISC40 sensor in Fig 19: Pipe/wall mounting kit flow fitting Note: The ISC40 sensor type -TG does not fit in the model ISC40FF-S. 10.3 Dimensions ISC40FF Option L1 /FF1...
  • Page 28: Modelcode Isc40Ff

    Option L1 Direction of FLOW /FS1 269-280 (10.6-11.0) 65 (2.56) /FS2 279-290 (11.0-11.4) 85 (3.35) /FS3 269-280 (10.6-11.0) 60.5 (2.38) /FS4 279-290 (11.0-11.4) 79.2 (3.12) Ø 54 (2.12) 120 (4.72) Fig 21: Flow fitting ISC40FF-S 10.4 Modelcode ISC40FF Model Suffix Option Description ISC40FF...
  • Page 29: Spareparts Isc40Ff

    10.5 Spareparts ISC40FF Options ISC40FF Flow fitting, flange adapters Part no. Description Process connection Material O-ring(s) K1521AL /FF1 DN15 PN10 PVDF Viton K1521AP /FF2 DN25 PN10 PVDF Viton K1521AE /FF3 ANSI ½” – 150 lbs PVDF Viton K1521AH /FF4 ANSI 1” – 150 lbs PVDF Viton K1521AM /FP1...
  • Page 30: Isc40Fd - Immersion Fitting

    11 ISC40FD – IMMERSION FITTING 11.1 General Specifications 11.1.1 Materials Wetted parts without options Model ISC40FD-S : AISI 316 SS, Viton Model ISC40FD-V : C-PVC, Viton Flange ISC40FD-S-**-SF* : AISI 316 SS Non-wetted parts Tube pigtail : plastic 11.1.2 Operating range Temperature Model ISC40FD-S : Max.
  • Page 31 ISC40FD-V ISC40FD-S Guide pipe Ø 48,6 x 600 (Ø1.9 x 24") Min: 110 (4.35") Max: 515 (20") O-Ring 44.17 x 1.78 O-Ring 39.35 x 2.62 2-inch Stanchion (Vertical) O-Ring 26.57 x 3.53 Fig 22: Installation of ISC40 sensor Fig 23: Installation on stanchion in immersion fitting.
  • Page 32: Dimensions Isc40Fd

    Alternative ways of mounting the immersion fitting are: • Guide pipe To facilitate this type of mounting the immersion fitting has a larger diameter at the top. The user supplied guide pipe (internal diameter > 52 mm) is fitted to the wall or mounting rail.
  • Page 33: Modelcode Isc40Fd

    11.4 Modelcode ISC40FD Model code Suffix code Option code Description ISC40FD Immersion fitting Material AISI 316 Stainless steel PVC-C Insertion length Between 05 to 20 Example: 05 = 0.5 m / 20 = 2.0 m Flange -NFL No flange -SFA AISI316 SS 2”...
  • Page 34: Chemical Compatibility Chart

    O O O O O O Note: Information in this list is based on our general experience and literature data and given in good faith. However Yokogawa is unable to accept responsobility for claims related to this information. IM 12D08J02-E-E...
  • Page 35 IM 12D08J02-E-E...
  • Page 36 Yokogawa has an extensive sales and YOKOGAWA ELECTRIC CORPORATION YOKOGAWA ELECTRIC ASIA Pte. LTD. 5 Bedok South Road World Headquarters distribution network. 9-32, Nakacho 2-chome, Musashino-shi Singapore 469270 Tokyo 180-8750 Singapore Please refer to the European website www.yokogawa.com/sg Japan www.yokogawa.com (www.yokogawa.com/eu) to contact your...

Table of Contents