Advertisement

Quick Links

Instruction
Manual
Models FD30 / PB30
Fittings for the DO30G
IM 12J06K04-E-E
6th edition

Advertisement

Table of Contents
loading

Summary of Contents for YOKOGAWA FD30

  • Page 1 Instruction Models FD30 / PB30 Manual Fittings for the DO30G IM 12J06K04-E-E 6th edition...
  • Page 2: Table Of Contents

    1. Introduction ......................3 2. Unpacking and checking ..................3 3. Warranty and service ....................3 4. Genral specifications ....................4 4-1. Model FD30 Immersion Fitting ................4 4.2 Model PB30 Floating Ball Fitting ................5 Installation ........................6 5.1 General........................6 5.2 Assembling and mounting of the immersion fitting..........6 5.2.1 Checklist ....................6...
  • Page 3: Introduction

    1. Introduction To support the Yokogawa Model DO30G sensor for measuring dissolved Oxygen, two types of sensor holders are offered; an immersion fitting (Model FD30V27) and a floating ball fitting (Model PB30). Both fittings are intended for mounting the DO30G in a tank, measuring either on the surface or sub-surface of the process liquid.
  • Page 4: Genral Specifications

    The shipping address where to the goods have to be returned is specified on the original sales order or on the back page of this manual. 4. General specifications 4.1 Model FD30 Immersion Fitting Wetted materials - Body & flange...
  • Page 5: Model Pb30 Floating Ball Fitting

    Model and Suffix code Immersion Fitting Model Suffix Option Description FD30V27 Immersion Fitting PVC Immersion Pipe supplied by user length Between 0.5 and 2.0 (in steps of 0.1m) Example: 06 = 0.6 m Mounting No flange flange DIN DN50 PN10 ANSI 2”...
  • Page 6: Installation

    5. Installation 5.1 General It is important to have the point of measurement in a location that is representing the process composition. Check whether the specifications of the sensor/fitting/holder fulfill the maximum occurring process conditions. Install the fitting in a place where it can be maintained and calibrated safely and easily.
  • Page 7: Preparing The Immersion Fitting

    5.2.2 Preparing the immersion fitting Prepare the immersion fitting by unscrewing the screw piece (4) from the bottom of the fitting. Remove the stopper (7), which can be discarded, and the conductive bushing (5). The screw piece will be used to fix the DO sensor in the fitting.
  • Page 8: Installing The Floating Ball Fitting

    5.2.4 Mounting the sensor cleaner (/JC option) • After fixing the sensor with the cleaning cage it is possible to screw the cleaner tubing on the cage using the swage lock coupling material provided with the cleaner. • It is recommended to fix the tubing to the holder using tie wraps.
  • Page 9: Spare Parts

    Figure 4 6. Spare parts 6.1 Model FD30V27 Immersion Fitting Item Description Spare part no 5 x O-ring K1500FX 5 x O-ring K1500FY 6.2 Sensor Cleaner (/JC option) Item Description Spare part no 5 x O-ring 37.7 X 2.6 for mounting sensor in fitting K1500FY Holder Nozzle parts...
  • Page 10: Dimensions

    Dimensions Figure 5: FD30 Figure 6: FD30 with flange 18 (0.70) 78 (3.07) 125 (4.92) 168 (6.60) 18 (0.70) Ø 100 (3.94) Ø 40 (1.57) Ø 40 (1.57) Ø 50 (1.97) Ø 50 (1.97) IM 12J06K04-E-E...
  • Page 11 Figure 7: Option /MS1 Guide pipe Fill piece Ø 48,6 x 600 (Ø1.9 x 24") Min: 110 (4.35") Max: 515 (20") 2-inch Stanchion (Vertical) Figure 8: PB30 rail mounting 2570 ± 50 (100 ± 2) Ø 50 (1.97) pipe/rail mounting DN 50 (2.00) 230 (9.06) IM 12J06K04-E-E...
  • Page 12 European website Fax (81)-422-55-1202 E-mail: info@yca.com (www.yokogawa.com/eu) E-mail: webinfo@mls.yokogawa.co.jp www.yca.com www.yokogawa.com to contact your nearest representative. YOKOGAWA ELECTRIC ASIA Pte. Ltd. YOKOGAWA EUROPE B.V. 5 Bedok South Road Euroweg 2 Singapore 469270 3825 HD AMERSFOORT Singapore The Netherlands Tel. (65)-241-9933 Tel.

This manual is also suitable for:

Pb30

Table of Contents