YOKOGAWA Sushi Sensor XS530 User Manual page 15

Pressure measurement module
Table of Contents

Advertisement

2.7.2
IECEx Intrinsic Safety
Technical data:
• Certificate number: IECEx DEK 19.0027X
• Applicable standards: I EC 60079-0 Ed. 7.0 (2017),
• Ex marking: Ex ib IIC T4 Gb
• Ambient temperature: −40 to 60°C (–40 to 140°F)
• Process temperature: −40 to 100°C (–40 to 212°F)
• Electrical parameters:
Connector
Ui = 6.88 V, Ii = 1.54 A, Pi = 0.3 W, Ci = 3.9 μF, Li = 0 μH
• Enclosure: IP66/IP67 in accordance with only IEC 60529 when combined
with XS110A.
Certification information:
WARNING
A modification of the equipment would no longer comply with the
construction described in the certificate documentation.
Control drawing:
Uo ≤ Ui
Io ≤ Ii
Po ≤ Pi
Co ≥ Ci
Lo ≥ Li
IS Apparatus
Note:
1. If earthing of the Case (metallic part of the enclosure) is not ensured by
installation, apply conductive connection between the Case and the earth
point (or the equipotential bonding system).
IEC 60079-11 Ed. 6.0 (2011)
Hazardous Location
Connector
Ui: 6.88 V
Ii: 1.54 A
20
Pi: 0.3 W
Ci: 3.9 µF
Li: 0 µH
Pressure measurement module
<Notes on Handling>
XS530
F03.ai
2-6
IM 01W06F01-01EN