4.2.1
Mounting with XS110A
This section describes the installation procedures of XS110A to the XS530.
Before installation, check that the battery is built-in the XS110A.
Refer to Section 4.1 "Notes on installation" before mounting with XS110A.
 Align and insert the XS110A and XS530 connectors.
 Fit the XS110A into the XS530.
 Turn the XS110A fixing screws attached to the XS530 one by one to the
left, pull them out, and turn them 1/4 left from where they come into contact
to tighten them. At this time, check that the head of the XS110A fixing screw
is flush with the surface of the XS110A.
 Turn the XS110A measuring module fixing screw to the right. At this time,
tighten the screws with a torque of 1.4 N·m.
After the XS110A has been installed, the settings can be made using the Sushi
Sensor App. First, make sure that the Sushi Sensor has the latest firmware. To
remove, follow the same procedure as for mounting. If it is difficult to pushup
the XS110A and XS530 by hand, you can easily pushup them by pushing a flat
blade screwdriver in between the grooves of the XS110A and XS530.
Figure 4-1 Mounting with XS110A
<Installation>
F05.ai
4-3
IM 01W06F01-01EN