2.2

Transportation

To prevent damage while in transit, leave the XS530 in the original shipping
container until it reaches the installation site.
2.3

Storage

When an extended storage period is expected, observe the following
precautions.
(1) Choose a storage location that satisfies the following requirements.
• A location that is not exposed to rain or water.
• A location subject to a minimum of vibration or impact.
• The following temperature and humidity range are recommended.
- Temperature: –40 to 85°C (–40 to 185°F)
- Humidity: 0 to 100% RH (non-condensation)
(2) When storing the XS530, repack it carefully in the packaging that it was
originally shipped with.
(3) If the XS530 has been used, thoroughly clean the chambers inside the
cover flanges, so that there is no process fluid remaining inside.
NOTE
When storing this product over a long period, it is recommended to put
the instrument in OFF Mode to conserve the battery. For details on how to
switch to OFF Mode, refer to "IM 01W06C01-01EN".
2.4

Selecting the Installation Location

This product is designed to withstand severe environmental conditions.
However, to ensure that it will provide years of stable and accurate
performance, take the following precautions when selecting the installation
location.