Operating the 280-DS
Bath Temperature Measurement
The Bath Temperature Measurement page displays the reading of the 280-DS
temperature probe and the bath temperature reading of the dissolution
apparatus, if it is online.
1 If the dissolution apparatus is offline, then you must enter the bath
NOTE
Figure 48.
2 After 30 seconds, the reading stabilizes. Click Accept to continue to the next
70
temperature as reported on its display. Although a successful test requires
reaching the target temperature, you can accept the result at anytime.
If applicable, do not remove the bath temperature probe of the
dissolution apparatus from the water bath for this measurement.
Instead, remove one of the rear dissolution vessels for access to
the water bath.
Bath Temperature Measurement
screen.