Operating the 280-DS
3 Slowly release the vessel diameter arm. Push down until the underside of the
Figure 34.
4 Lower the paddle or basket to the operating position.
5 Check the AS FOUND values to ensure the shaft has been detected and the
6 Click Run to move on to the first in-vessel test.
WARNING
56
top of the Vessel Module touches the top of the vessel collar. Ensure that the
Vessel Module height arm extension is touching the bottom of the vessel.
VM Resting at the Bottom of the Vessel
VM is properly seated. The AS FOUND values are displayed to ensure proper
placement. The AS FOUND values will be stored and displayed on the final
report when you select Run.
The Vessel Module contains several components that are spring
loaded. Use care to avoid pinching or damaging the vessel when
releasing the diameter arm.