Agilent Technologies 280-DS Operator's Manual page 39

Mechanical qualification system
Table of Contents

Advertisement

4
Operating the 280-DS
Method Editor Specification Tab
1 Click the Specification tab. The Specification page provides the list of
2 Edit the default specifications by clicking on each one and modifying the
Figure 17.
NOTE
3 You can click OK or Cancel on any page of this sequence. If you have changed
280-DS Operator's Manual
available tests and fields for entering specification limits and or nominal
values. Many default specifications (e.g., USP, FDA, ASTM) are available for
your convenience. The pre-entered options will change and/or will be grayed
out depending on your configuration.
values. If any adjustment is made to the pre-loaded specifications, Custom
will display as the Compliance Specification. Place a check in the n/a box to
exclude that test from the method.
Specification Tab
Separate additional RPMs in the Rotational Speed box by using a
comma, space, or semicolon.
the method, you will be asked for a reason for the change. After you enter a
reason for the change, the method data, new or modified, is saved to the
database. If you created a new method, the method version will be indicated
by the number 1. If you modified a method, its version will be increased by 1
39