Agilent Technologies 280-DS Operator's Manual page 50

Mechanical qualification system
Table of Contents

Advertisement

4
Operating the 280-DS
Vessel Module Vessel Centering Arm Placement
C A U T I O N
If any of the shaft/vessel alignment tests are required then the software prompts
you to zero the encoder of the diameter arm of the Vessel Module.
1 Hold the Vessel Module in a way that does not restrict the movement of the
C A U T I O N
2 Click Zero to continue the test.
NOTE
50
Ensure the manifold and evaporation covers on the dissolution
apparatus are removed prior to using the Vessel Module. Refer
to the applicable section in the documentation of your
dissolution apparatus for removal instructions.
diameter arm
(Figure 28
Do not place a finger in the cavity in the bottom of the Vessel
Module. Doing so can contaminate the lenses inside and cause
invalid readings.
Prior to measuring parameters within the dissolution vessel,
please ensure that the vessel orientation at each active location is
properly established.
on page 51).
280-DS Operator's Manual