Agilent Technologies 280-DS Operator's Manual page 59

Mechanical qualification system
Table of Contents

Advertisement

4
Operating the 280-DS
Vessel / Shaft Verticality and Vessel Centering
The following tests will be executed simultaneously on this screen:
Figure 37.
NOTE
1 Once satisfactory results have been achieved, you can document the Vessel
NOTE
2 Click Accept to continue to the next screen.
280-DS Operator's Manual
Vessel Verticality - 2 points, 90 deg. apart
Shaft Verticality - 2 points, 90 deg. apart
Vessel / Shaft Centering (Upper and Lower) - Upper and Lower vessel
position
Verticality and Centering Measurement
You can make adjustments to the dissolution apparatus at any
time prior to clicking Accept to optimize the performance of the
instrument.
Orientation in the open text field. The description you enter will be displayed
on the report and stored for future reference.
Comments entered under the Common tab apply to all of the tests
while
comments entered under the Vessel Verticality, Shaft Verticality, or
Vessel/Shaft Centering tabs apply only to the individual test.
59

Advertisement

Table of Contents
loading

This manual is also suitable for:

G7980aa

Table of Contents