4
Operating the 280-DS
Vessel / Shaft Verticality and Vessel Centering
The following tests will be executed simultaneously on this screen:
Figure 37.
NOTE
1 Once satisfactory results have been achieved, you can document the Vessel
NOTE
2 Click Accept to continue to the next screen.
280-DS Operator's Manual
•
Vessel Verticality - 2 points, 90 deg. apart
•
Shaft Verticality - 2 points, 90 deg. apart
•
Vessel / Shaft Centering (Upper and Lower) - Upper and Lower vessel
position
Verticality and Centering Measurement
You can make adjustments to the dissolution apparatus at any
time prior to clicking Accept to optimize the performance of the
instrument.
Orientation in the open text field. The description you enter will be displayed
on the report and stored for future reference.
Comments entered under the Common tab apply to all of the tests
while
comments entered under the Vessel Verticality, Shaft Verticality, or
Vessel/Shaft Centering tabs apply only to the individual test.
59