Configuring the UTREL-16 for logging
 
A tertiary upper alarm cannot be entered with a lower threshold value than the
l
secondary alarm. Threshold values can, however, be equal and combined with
different time-to-activation values. For example, you can enter a primary upper
alarm with an -70.0 °C threshold and 10 accumulative readings, plus an alarm with
an -70.0 °C threshold and 5 consecutive readings. In this instance the alarm will be
triggered, if either 10 readings in total are above -70.0 °C, or 5 consecutive readings
above -70.0 °C have occurred.
If an alarm is triggered, you will be able to see this on the PDF, in the software and on the
display.
The section about
Reviewing Statistics and Alarms on page 26
is represented and reviewed on the display.

Finalizing the configuration

Click
Configure
to upload the configuration data to the UTREL-16.
When the configuration is complete, unplug the UTREL-16 from the USB socket and
replace the protective cap.
If you wish to configure more UTREL-16 units with the same configuration, insert the next
loggers into USB sockets, wait until they are ready for configuration and click
Configure. Alternatively, you can use the
the same settings.
You can upload the configuration to a UTREL-16 logger as often as required.
UTREL-16 User Guide
Profile
function to configure multiple units with
shows examples of how this
Repeat