Operating the Logger
 
Releasing the button within the 5-second period after the initial press switches to
l
the first review screen.
A min/max reset mark will be recorded in the logged data. The logger will now track new
min/max values, with the first update shown as soon as the next reading is taken.
You will be able to see in the chart in both LogTag
min/max values were cleared, but you will not be able to review previous min/max values
on screen once they have been cleared.
Clearing an Alarm
A UTREL-16 can be configured so you can clear an alarm from the display. To do this,
press and hold the
After 2 seconds the alarm will be cleared and the
the button for a further 2 seconds, the
An inspection mark will be recorded against the next temperature reading taken, whether
you clear the alarm or not.
An alarm cannot be cleared once the UTREL-16 has stopped recording. Alarms
resulting from an accumulation of readings can also not be cleared.
UTREL-16 User Guide
button.
START
/Mark
®
Analyzer and the PDF report when the
will change to a
will show again and the alarm will not be cleared.